Frequently Asked Questions

Available Resources

 

What microscopy resources are available for use?

We have two TEMs appropriate for single-particle imaging and, occasionally, related microscopy projects — a Thermo Fisher Tundra (100 kV) and a JEOL 1400Plus (80 – 120 kV). The Tundra is maintained by and for the Chemical Biology Program at Dana-Farber Cancer Institute, whereas the JEOL 1400Plus is maintained by and for the Dana-Farber Cancer Institute Department of Cancer Biology; access is prioritized accordingly.

Do you have resources for plunging or clipping grids or provide assistance with higher-resolution data collection?

We provide resources and training for clipping grids and screening them on the Tundra. We do not provide a plunger for making grids or higher-voltage microscopes for high-resolution data collection. Both of these are available at the HMS cryo-EM facility and we can help connect you as needed. Note that an approved project proposal is necessary in order to utilize HMS microscopes.

Do you have resources for computation or storage? 

We do not provide storage space or computational resources, with the exception of fully-assisted processing, in which case we will provide resources for processing but you will be responsible for long-term storage of your data set and its associated processing. Prior to starting on the Tundra, you must have adequate space (~2-3 TB per data set) to store movies and processing data. We can advise you on how to access HMS / Dana-Farber Cancer Institute storage and computational resources or purchasing a local workstation; some resources to consider are discussed here.

Will you prepare grids / screen, collect, and process data for me?

The Tundra is a training instrument, not a service core; we prioritize training users to independence and expect users to be present and active during all parts of the process, even for a fully assisted project. You may want to contact our neighbors in the Structural & Chemical Biology Center if you are interested in core crystallography services. We have an hourly consulting rate if you need assistance with getting started with the processing workflow or have particularly difficult processing questions, but we encourage new labs to collaborate with established structural biology labs until they are more familiar with the workflow and resources required for success in cryo-EM.

Tundra Location and Access

 

Where is the Tundra located?

The Tundra is located in Longwood Center, room 3303.

Who can use the Tundra?

The Tundra is an instrument maintained by and for the chemical biology program at Dana-Farber. We are keen to enable broader Dana-Farber usage, but members of the program have scheduling priority. Researchers outside of Dana-Farber are not currently able to use the Tundra.

Training and Cost

 

What does the training process involve?

Clipping training takes approximately 2 hours per session and occurs over approximately three sessions with increasing independence. These sessions occur on a separate day than Tundra training.

Tundra training usually encompasses three to four all-day sessions, some of which may include overnight data collections. Once sufficient mastery and independence have been demonstrated, users can transition to independent use of the instrument.

Training for negative stain grid preparation takes approximately 2 hours. Training for JEOL use occurs over approximately three sessions with increasing independence. Each session is approximately 3 hours in length. 

Can I get trained with a friend?

Because the training process is hands-on, we require that each person undergo the full training process independently. Others are welcome to observe during any part of the process! 

Training for negative stain grid preparation and initial JEOL training can occur with one other person as long as this is arranged in advance.

What is the cost to use the Tundra?

The current pricing structure is on the pricing page; services must be requested via iLabs and will be charged to the project you provide. The pricing of the instrument will be subject to change as we are striving to offer the lowest possible rates, while recovering costs for service contracts and staff; any changes to pricing will be communicated prior to implementation. 

Tundra FAQ

 

How many samples can be screened in a day?

A fully trained user can screen approximately eight grids per day, or screen approximately six grids and set up one overnight data collection. Initial training will usually cover this number of grids in two successive days.  

What is the maximum resolution a data set collected on the Tundra can reach after processing?

The Tundra can reach slightly under 3 A resolution on a high-symmetry test protein (apoferritin). Users with high-quality grids often reach 4 - 5 A resolution from a small (~1k movies) data set on the Tundra before taking a similar grid to the Krios. 

What software do you use for screening and collection?

We currently screen and collect using EPU (Thermo Fisher). If you have an application that requires SerialEM or another program, contact Talya to discuss further.   

What is the difference between the Tundra and a higher-voltage microscope (e.g., a Krios or Arctica)?

The Tundra accelerates electrons at 100 kV, which is lower than the Arctica or Glacios (200 kV) or Krios (300 kV). This leads to higher-contrast images but a loss of more high-resolution information. 100 kV electrons also require a specialized detector and objective lens system in order to capture adequate information from the sample. In addition, the Tundra is designed to be an entry-level microscope and has various adaptations to make it more user-friendly and cost-effective, including a unique grid loading system, lack of an autoloader, and automation of some alignments. 

Can the Tundra do negative stain / tilted data sets / microED / {insert microscopy technique here}?

The Tundra is able to take data sets on negative stain grids as long as they are not stained with radioactive isotopes (i.e. uranyl compounds); we can provide alternative stains (NanoW, Uranyless) as needed after consultation. Routine imaging of negative stain grids should occur on the JEOL 1400Plus, which is maintained by the Cancer Biology department and is also located on the third floor of Longwood Center. In addition, the Tundra stage is unable to tilt due to its specialized objective lens system with a small pole piece gap, and so we are unable to take tilted data sets or perform microED.

The Tundra is an instrument optimized for cryo-EM, and so alternative applications will generally be better suited for other microscopes; however, if you have a nonstandard project that you think could utilize the Tundra, please reach out and we can discuss further.

Data Retention and Privacy

 

How long will my data be stored on the facility server?

We encourage you to transfer your data immediately after collection, we guarantee that your data will be accessible on the server for 14 days after collection. Data will never remain on the server for more than six weeks and can be deleted sooner upon user request. 

How private is my data?

Your data can technically be accessed by other Dana-Farber Cancer Institute Tundra users and SBGrid administrators during the time it is stored on our server, but not by anyone outside of the Dana-Farber Cancer Institute Tundra user base. If you have concerns about data privacy, please contact Talya. 

Labs without Structural Biology Expertise

If you are new to structural biology, we highly encourage you to seek out an experienced structural biology collaborator to better understand the cryo-EM process and resources required prior to working independently. However, we are happy to train new structural biologists as long as the lab demonstrates an understanding of and commitment to the time, personnel, financial, and computational resources needed for a cryo-EM project to be successful.  

Is my favorite protein / complex a good candidate for cryo-EM?

Generally, mostly-structured and inflexible proteins or stable complexes that are between 100 kDa and 1 MDa, can be concentrated in a mostly-pure form to 2-5 uM, and are stable in buffers without high concentrations of glycerol, salts, or DMSO are promising candidates for cryo-EM. However, even promising candidates can be difficult or impossible to solve using cryo-EM because of their behavior during the vitrification process. Take a look at the teaching tools page for some helpful posters on this front!

How soon will I get a high-resolution structure?

Many factors influence if it is possible and the time it takes to solve a high-resolution cryo-EM structure, including protein size, protein flexibility, buffer composition, complex stability, air/water interface effects, orientation bias, user expertise, and computational resource availability. Some targets can be solved relatively quickly (months) by an experienced user, and some difficult targets are ultimately intractable to solve by cryo-EM after years of effort by experienced structural biologists.

Where to Start / Other Questions

Contact Talya (talya_levitz@dfci.harvard.edu) and I will be happy to discuss your project or questions further.